2014

Glucose biosensor based on multisegment nanowires exhibiting reversible magnetic control.

Talanta Gerola GP, Takahashi GS, Perez GG, Recco LC, Pedrosa VA
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We describe the amperometric detection of glucose using oriented nanowires with magnetic switching of the bioelectrochemical process. The fabrication process of the nanowires was prepared through controlled nucleation and growth during a stepwise electrochemical deposition, and it was characterized using scanning electron microscopy. Cyclic voltammetry and amperometry were used to study the magnetoswitchable property; this control was accomplished by changing the surface orientation of nanowires. Under the optimal condition, the amperometric response was also linear up to a glucose concentration of 0.1-16.0 mmol L(-1) with a sensitivity of 81 μA mM(-1). The detection limit was estimated for 4.8×10(-8) mol L(-1), defined from a signal/noise ratio of 3. It also exhibits good reproducibility and high selectivity with insignificant interference from ascorbic acid, acetoaminophen, and uric acid. The resulting biosensor was applied to detect the blood sugar in human serum samples without any pretreatment, and the results were comparatively in agreement with the clinical assay.

关键词