2014

Construction of an uricase nanoparticles modified au electrode for amperometric determination of uric acid.

Applied biochemistry and biotechnology Chauhan N, Kumar A, Pundir CS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A method is described for preparation of uricase nanoparticles (100 nm in size) and their direct immobilization onto the Au electrode. The enzyme electrode along with Ag/AgCl as reference and Pt as auxiliary electrode were connected through potentiostat/galvanostat to construct an amperometric uric acid biosensor. The enzyme electrode was characterized by scanning electron microscopy (SEM), Fourier transform infrared (FTIR) spectroscopy, and electrochemical impedance spectroscopy (EIS). The enzyme electrode detected uric acid level as low as 5.0 μM at a signal-to-noise ratio of 3, within 7 s at pH 8.5 and 40 °C. The biosensor showed a linear working range, 0.005 to 0.8 mM for uric acid with a sensitivity of 0.03 mA μM(-1) cm(-2). The biosensor was evaluated. The biosensor lost only 15 % of its initial activity over a period of 7 months, when stored at 4 °C. The fabricated biosensor was successfully employed for determination of uric acid in human serum and urine.

关键词