Integrated catheter system for continuous glucose measurement and simultaneous insulin infusion.

Biosensors & bioelectronics Nacht B, Larndorfer C, Sax S, Borisov SM, Hajnsek M, Sinner F, List-Kratochvil EJ, Klimant I
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A new measurement system enables combination of continuous glucose monitoring (CGM) and insulin infusion. A sensor system comprising an optical glucose biosensor and an optical oxygen sensor is integrated into the insulin infusion catheter of an insulin pump. Both sensors rely on near infrared (NIR) phosphorescent porphyrin dyes, wherefore the signals can be read out transcutaneous and non-invasively with a custom-built phase fluorometer measurement module. The spectral properties of the indicator dyes and the optical setup of the measurement module were optimized to enable independent read-out in two channels. Dynamic ranges from 0 mmHg to 160 mmHg oxygen and 0mg/dL to 360 mg/dL glucose (LOD 2mg/dL) are covered by the oxygen and the glucose sensor, respectively. In-vivo measurements in pigs demonstrate good correlation of reference blood glucose levels and glucose values obtained with the presented sensor system. The evaluation of the clinical accuracy of the system with Clarke Error Grid Analysis showed similar results to CGM-devices currently on the market.