2015

MWCNT-cysteamine-Nafion modified gold electrode based on myoglobin for determination of hydrogen peroxide and nitrite.

Bioelectrochemistry (Amsterdam, Netherlands) Canbay E, Şahin B, Kıran M, Akyilmaz E
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this work, a novel amperometric biosensor of hydrogen peroxide (H2O2) was developed based on the immobilization of myoglobin (Mb) on the surface of the multi-walled carbon nanotube (MWCNT) -Nafion-cysteamine (CA) modified gold electrode (Au) and its electrocatalytic activity was used for the determination of nitrite (NO2(-)). In the optimization studies, the best MWCNT and myoglobin amount were investigated. It was discovered at the experiments for the optimization of the working conditions that the buffer at this study as 50.0mM, pH7.0 phosphate buffer (PBS) and working temperature as 30°C for the H2O2 biosensor. It was determined at the characterization studies on the biosensor that linear results are obtained between the ranges of 0.1μM to 70.0μM for H2O2 concentration and 1-250μM for NO2(-). The reproducibility of the biosensor was determined both H2O2 and nitrite. From the experiments, average value, standard deviation (SD) and coefficients of variation (CV%) were calculated to be 10.02±0.43μM, and 4.29% for 10.0μM H2O2 (n=6) and 52.0±2.1μM, and 3.89% for 50.0μM nitrite (n=8), respectively. At the same time the sample was analyzed for NO2(-) in drinking and mineral waters.

关键词