DNA-based biosensors for Hg(2+) determination by polythymine-methylene blue modified electrodes.

Biosensors & bioelectronics Tortolini C, Bollella P, Antonelli ML, Antiochia R, Mazzei F, Favero G
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this work we have developed a new electrochemical DNA-based biosensor for the selective determination of the Hg(2+) ion by the use of different electrodes modified with polythymine, bearing methylene blue, as redox probe, in 3' position. The determination of Hg(2+) can be employed with an excellent degree of selectivity by the use of DNA biosensors through the formation of the complex Thymine-Hg-Thymine (T-Hg-T): in fact, Hg(2+) tends to bind two thymines, generating a T-Hg-T complex with a formation constant higher than that one of the coupling Adenine-Thymine, which can be employed for a selective, fast and cost-effective Hg(2+) detection. The presence of the Hg(2+) in solution leads to the formation of T-Hg-T complex thus causing the "hairpin-like" folding of oligonucleotide, leading to an improved electronic exchange of methylene blue with the electrode surface due to the reduced distance and thus to an increase of the faradic current which is detected by means of square wave voltammetry (SWV). To test the feasibility of this kind of biosensor to be applied to the analysis of Hg(2+) we have developed several biosensors configuration by modifying the electrochemical sensor transducer: (a) Au electrode; (b) Au screen-printed electrode (SPE). The proposed system, allows the determination of Hg(2+) in the range 0.2-100 nM (0.05-20 ppb), with a sensitivity 0.327 µA/nM, LOD 0.1 nM (0.02 ppb), LOQ 0.2 nM (0.05 ppb) and RSD ≤4.3% when Au electrode is used as electrochemical transducer; on the other hand, in the case of Au SPE the linear range is 0.2-50 nM (0.05-10 ppb), with a sensitivity 0.285 µA/nM, while LOD and LOQ are the same as previously and RSD is ≤3.8%. This enabled the detection of mercury in real samples (waters and fishes) with good accuracy (recoveries 92-101% on waters and 92-107% on fishes, respectively) and reproducibility (RSD ≤9.6% for measurements on waters and ≤8.8% on fishes, respectively).

关键词