组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Reduced graphene oxide (rGO) has been fabricated into a microelectrode array (MEA) using a modified nanoimprint lithography (NIL) technique. Through a modified NIL process, the rGO MEA was fabricated by a self-alignment of conducting Indium Tin Oxide (ITO) and rGO layer without etching of the rGO layer. The rGO MEA consists of an array of 10μm circular disks and microelectrode signature has been found at a pitch spacing of 60μm. The rGO MEA shows a sensitivity of 1.91nAμm(-1) to dopamine (DA) without the use of mediators or functionalization of the reduced graphene oxide (rGO) active layer. The performance of rGO MEA remains stable when tested under highly resistive media using a continuous flow set up, as well as when subjecting it to mechanical stress. The successful demonstration of NIL for fabricating rGO microelectrodes on flexible substrate presents a route for the large scale fabrication of highly sensitive, flexible and thin biosensing platform.