2014

Electrochemical biochip for applications to wireless and batteryless monitoring of free-moving mice.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference Baj-Rossi C, De Micheli G, Carrara S
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A multi-sensing platform for applications in wireless and batteryless monitoring of free-moving small animals is presented in this paper. The proposed platform hosts six sensors: four biosensors for sensing of both disease biomarkers and therapeutic compounds, and two further sensors (T and pH) for biosensor calibration. Electrodeposition of Multi-Walled Carbon Nanotubes (MWCNTs) and the subsequent function-alization with proper enzymes is used to assure sensitivity and specificity in electrochemical biosensing. The realized sensors are demonstrated to be capable of measuring several parameters: lactate with a sensitivity of 77±26 μA/mM· cm(2) and a limit of detection (LOD) of 4±1 μM; glucose with a sensitivity of 63±15 μA/mM· cm(2) and a LOD of 8±2 μM; Etoposide (a well known anti-cancer agent) with a sensitivity of 0.15±0.04 mA/mM· cm(2) and a LOD of 4±1 μM; Open Circuit Potential (OCP) measurements are used on a Pt/IrOx junction to sense pH with a sensitivity of around -75±5mV/pH; while a Pt resistive thermal device is used to measure physiological temperature-range with an average sensitivity of 0.108±0.001 kΩ/°C.

关键词