A near infrared holographic glucose sensor.

Biosensors & bioelectronics Vezouviou E, Lowe CR
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Real-time glucose monitoring has been beneficial in reducing health complications associated with diabetes as well as a decrease in mortality. This report describes a novel holographic platform, fabricated via laser ablation on chitosan hydrogel with gold nanoparticles with a replaying in visible and near IR. The sensor responded with a 12 nm and 7 nm shift in wavelength at glucose concentrations in the 0-70 mM range and in the visible and near IR, respectively, at pH 7.4 and an ionic strength of 154 mM. The sensor did not respond to potential interferences found in the interstitial fluid, such as fructose, vitamin C and lactate, at their respective normal concentrations and was stable to fluctuations in temperature, pH and ionic strength. The characteristics of this sensor suggests that it may be applicable for use as an implanted device for the real time monitoring of glucose concentrations in the interstitial fluid using near IR as the interrogating medium.