2015

Direct determination of a small-molecule drug, valproic Acid, by an electrically-detected microcantilever biosensor for personalized diagnostics.

Biosensors Huang LS, Gunawan C, Yen YK, Chang KF
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Direct, small-molecule determination of the antiepileptic drug, valproic acid, was investigated by a label-free, nanomechanical biosensor. Valproic acid has long been used as an antiepileptic medication, which is administered through therapeutic drug monitoring and has a narrow therapeutic dosage range of 50-100 μg·mL-1 in blood or serum. Unlike labeled and clinically-used measurement techniques, the label-free, electrical detection microcantilever biosensor can be miniaturized and simplified for use in portable or hand-held point-of-care platforms or personal diagnostic tools. A micromachined microcantilever sensor was packaged into the micro-channel of a fluidic system. The measurement of the antiepileptic drug, valproic acid, in phosphate-buffered saline and serum used a single free-standing, piezoresistive microcantilever biosensor in a thermally-controlled system. The measured surface stresses showed a profile over a concentration range of 50-500 μg·mL-1, which covered the clinically therapeutic range of 50-100 μg·mL-1. The estimated limit of detection (LOD) was calculated to be 45 μg·mL-1, and the binding affinity between the drug and the antibody was measured at around 90 ± 21 μg·mL-1. Lastly, the results of the proposed device showed a similar profile in valproic acid drug detection with those of the clinically-used fluorescence polarization immunoassay.

关键词