2015

New approach for natural products screening by real-time monitoring of hemoglobin hydrolysis using quartz crystal microbalance.

Analytica chimica acta Cornelio VE, Pedroso MM, Afonso AS, Fernandes JB, da Silva MF, Faria RC, Vieira PC
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The hemoglobin (Hb) released from erythrocytes is a primary nutritive component for many blood-feeding parasites. The aspartic protease cathepsin D is a hemoglobinase that is involved in the Hb degradation process and is considered an interesting target for chemotherapy intervention. However, traditional enzymatic assays for studying Hb degradation utilize spectrophotometric techniques, which do not allow real-time monitoring and can present serious interference problems. Herein, we describe a biosensor using simple approach for the real-time monitoring of Hb hydrolysis as well as an efficient screening method for natural products as enzymatic inhibitors using a quartz crystal microbalance (QCM) technique. Hemoglobin was anchored on the quartz crystal surface using mixed self-assembled monolayers. The addition of the enzyme caused a mass change (frequency shift) due to Hb hydrolysis, which was monitored in real time. From the frequency change patterns of the Hb-functionalized QCM, we evaluated the enzymatic reaction by determining the kinetic parameters of product formation (k(cat)). The QCM enzymatic assay using immobilized human Hb was shown to be an excellent approach for screening possible inhibitors in complex mixtures, opening up a new avenue for the discovery of novel inhibitors.

关键词