2015

Detection of ochratoxin A using molecular beacons and real-time PCR thermal cycler.

Toxins Sanzani SM, Reverberi M, Fanelli C, Ippolito A
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We developed a simple and cheap assay for quantitatively detecting ochratoxin A (OTA) in wine. A DNA aptamer available in literature was used as recognition probe in its molecular beacon form, i.e., with a fluorescence-quenching pair at the stem ends. Our aptabeacon could adopt a conformation allowing OTA binding, causing a fluorescence rise due to the increased distance between fluorophore and quencher. We used real-time PCR equipment for capturing the signal. With this assay, under optimized conditions, the entire process can be completed within 1 h. In addition, the proposed system exhibited a good selectivity for OTA against other mycotoxins (ochratoxin B and aflatoxin M1) and limited interference from aflatoxin B1 and patulin. A wide linear detection range (0.2-2000 µM) was achieved, with LOD = 13 nM, r = 0.9952, and R2 = 0.9904. The aptabeacon was also applied to detect OTA in red wine spiked with the same dilution series. A linear correlation with a LOD = 19 nM, r = 0.9843, and R2 = 0.9708 was observed, with recoveries in the range 63%-105%. Intra- and inter-day assays confirmed its reproducibility. The proposed biosensor, although still being finalized, might significantly facilitate the quantitative detection of OTA in wine samples, thus improving their quality control from a food safety perspective.

关键词