2015

Building an aptamer/graphene oxide FRET biosensor for one-step detection of bisphenol A.

ACS applied materials & interfaces Zhu Y, Cai Y, Xu L, Zheng L, Wang L, Qi B, Xu C
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Bisphenol A (BPA) is an important industrial chemical for polycarbonate (PC) and epoxy resins in paper and plastic industries. In our work, a kind of new method for detection of BPA was designed based on graphene oxide and anti-BPA aptamer. The graphene oxide can specifically adsorb and quench the fluorescence of fluorescently modified ssDNA probes. Meanwhile, the BPA can combine with anti-BPA optamer and switch its configuration to prevent the aptamer from adsorbing on the surface of graphene oxide (GO). Under different concentrations of BPA, based on the target-induced conformational change of anti-BPA aptamer and the interactions between the fluorescently modified anti-BPA aptamer (FAM-ssDNA) and GO, the experimental results show that the intensity of the fluorescence signal was changed. A low limit of detection of 0.05 ng/mL was obtained in the range 0.1-10 ng/mL. In addition, the specificity was outstanding among analogues of BPA. The recovery rate in actual water samples spiked with BPA can be 96.0% to 104.5%. The developed method was successfully used to determine BPA in actual water samples.

关键词