2015

Ultrasensitive Impedimetric Biosensor Fabricated by a New Immobilisation Technique for Parathyroid Hormone.

Applied biochemistry and biotechnology Özcan HM, Yildiz K, Çakar C, Aydin T, Asav E, Sağiroğlu A, Sezgintürk MK
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This paper presents a novel ultrasensitive and rapid impedimetric biosensor with new immobilisation materials for parathyroid hormone (PTH) with the aim to determine the PTH level in serum for the diagnosis and monitoring of parathyroid diseases such as hyperparathyroidism, adenoma, and thyroid cancer. The interaction between PTH and the biosensor was investigated with an electrochemical method. The biosensor was based on the gold electrode modified by mercaptohexanol (6-MHL). Anti-parathyroid hormone (anti-PTH) was covalently immobilised onto a self-assembled monolayer (SAM) by using epiclorhidrina (EPI) with ethanolamine (EA). The EPI-EA interaction represents the first use of these for the construction of biosensors in published reports. The immobilisation of the anti-PTH was monitored by electrochemical impedance spectroscopy, cyclic voltammetry and scanning electron microscopy (SEM) techniques. After the optimisation studies of immobilisation materials such as 6-MHL, EPI, EA and glutaraldehyde, linearity, repeatability and sensitivity of biosensor were evaluated as the performance of biosensor. PTH was detected within a linear range of 0.1-0.6 pg/ml, and the detection limit was 0.1 fg/ml. The specificity of the biosensor was also investigated. Finally, the described biosensor was used to detect the PTH levels in artificial serum samples.

关键词