2015

Feasible electrochemical biosensor based on plasma polymerization-assisted composite of polyacrylic acid and hollow TiO2 spheres for sensitively detecting lysozyme.

Biosensors & bioelectronics Zhang Z, Zhang S, He L, Peng D, Yan F, Wang M, Zhao J, Zhang H, Fang S
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A composite made of polyacrylic acid and hollow TiO2 spheres (TiO2@PPAA) was prepared by the plasma polymerization method and subsequently used as an electrode material for detecting lysozyme. The chemical structure, surface morphology, and electrochemical performance of the TiO2@PPAA composite were mainly affected by the plasma input power used during plasma polymerization. After optimizing plasma conditions, aptamer strands exhibited high adsorption affinity toward the surface of TiO2@PPAA composite via synergistic effects between TiO2 and PPAA. Electrochemical impedance spectroscopy results showed that the developed TiO2@PPAA aptasensor presents highly sensitive detection ability toward lysozyme; the limit of detection of the proposed aptasensor is 0.015 ng mL(-1) (1.04 pM) within the range of 0.05-100 ng mL(-1) in terms of 3σ value. The film further showed excellent selectivity toward lysozyme in the presence of interfering proteins, such as thrombin, bovine serum albumin, and immunoglobulin E. Thus, this aptasensing strategy might broaden the applications of plasma polymerized nanomaterials in the field of biomedical research and early clinical diagnosis.

关键词