2016

Intense charge transfer surface based on graphene and thymine-Hg(II)-thymine base pairs for detection of Hg(2.).

Biosensors & bioelectronics Li J, Lu L, Kang T, Cheng S
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this article, we developed an electrochemiluminescence (ECL) sensor with a high-intensity charge transfer interface for Hg(2+) detection based on Hg(II)-induced DNA hybridization. The sensor was fabricated by the following simple method. First, graphene oxide (GO) was electrochemically reduced onto a glassy carbon electrode through cyclic voltammetry. Then, amino-labeled double-stranded (ds)DNA was assembled on the electrode surface using 1-pyrenebutyric acid N-hydroxysuccinimide as a linker between GO and DNA. The other terminal of dsDNA, which was labeled with biotin, was linked to CdSe quantum dots via biotin-avidin interactions. Reduced graphene oxide has excellent electrical conductivity. dsDNA with T-Hg(II)-T base pairs exhibited more facile charge transfer. They both accelerate the electron transfer performance and sensitivity of the sensor. The increased ECL signals were logarithmically linear with the concentration of Hg(II) when Hg(2+) was present in the detection solution. The linear range of the sensor was 10(-11) to 10(-8)mol/L (R=0.9819) with a detection limit of 10(-11)mol/L. This biosensor exhibited satisfactory results when it was used to detect Hg(II) in real water samples. The biosensor with high-intense charge transfer performance is a prospect avenue to pursue more and more sensitive detection method.

关键词