2016

Preparation of hemoglobin (Hb) imprinted polymer by Hb catalyzed eATRP and its application in biosensor.

Biosensors & bioelectronics Sun Y, Du H, Lan Y, Wang W, Liang Y, Feng C, Yang M
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Molecularly imprinted polymer (MIP) was prepared on the surface of Au electrode by electrochemically mediated atom transfer radical polymerization (eATRP) with hemoglobin (Hb) both as catalyst and template molecule. Firstly, the condition for eATRP such as the potential, time and Hb concentration were selected and determined to be -0.51 V, 120 min and 20mg/mL, respectively. Further, the electrode modified with MIP (MIP/Au) was carefully examined by cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS) and scanning electron microscope (SEM). Finally, the MIP/Au electrode was used as a biosensor and successfully applied for the determination of Hb by differential pulse voltammetry (DPV) measurement. The results of experiments showed that the proposed biosensor displayed a broader linear range and a lower detection limit for Hb determination when it was compared to those Hb sensors based on MIP. The linear range was from 1.0 × 10(-10) to 1.0 × 10(1)mg/L with a detection limit of 7.8 × 10(-11)mg/L (S/N=3.3). In a word, the work of this paper established a useful way for the preparation and application of biosensor based on protein imprinted polymers.