组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
A magnetite and gold nanoparticle modified carbon paste electrode (CPE) was prepared for the immobilization of a thiol modified Hepatitis B virus (HBV) probe DNA and determination trace amount of target HBV DNA. Indeed, the sensing platform integrated two nanoparticles that had previously been employed individually in the DNA biosensors. The proposed DNA biosensor could measure target HBV DNA virus concentration with a low detection limit of 3.1 (±0.1)×10-(13)M, which was greatly lower than the detection limit reported with gold or magnetite nanoparticles alone. The change of interfacial charge transfer resistance (RCT) was confirmed the hybrid formation between probe and target HBV DNA. The RCT difference (before and after hybridization with the target HBV DNA) was in a linear relationship with the logarithm of complementary oligonucleotide concentrations in the range of 8.3 (±0.1)×10(-13) to 6.4 (±0.2)×10(-7)M. In addition, the novel methodology for specific DNA sequence detection was highly selective, repeatable, and reproducible. Finally, this work was successfully utilized for the sensitive and label free impedimetric determination of HBV target DNA in the urine and blood plasma samples.