2016

A novel amperometric biosensor for ß-triketone herbicides based on hydroxyphenylpyruvate dioxygenase inhibition: A case study for sulcotrione.

Talanta Rocaboy-Faquet E, Barthelmebs L, Calas-Blanchard C, Noguer T
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

An amperometric biosensor was designed for the determination of sulcotrione, a β-triketone herbicide, based on inhibition of hydroxyphenylpyruvate dioxygenase (HPPD), an enzyme allowing the oxidation of hydroxyphenylpyruvate (HPP) in homogentisic acid (HGA). HPPD was produced by cloning the hppd gene from Arabidopsis thaliana in E. coli, followed by overexpression and purification by nickel-histidine affinity. The electrochemical detection of HPPD activity was based on the electrochemical oxidation of HGA at +0.1 V vs. Ag/AgCl, using a poly(3,4-ethylenedioxythiophene) polystyrene sulfonate-modified screen-printed electrode. Assays were performed at 25°C in 0.1 M phosphate buffer pH 8 containing 0.1M KCl. The purified HPPD was shown to display a maximum velocity of 0.51 µM(HGA) min(-1), and an apparent K(M) of 22.6 µM for HPP. HPPD inhibition assays in presence of sulcotrione confirmed a competitive inhibition of HPPD, the calculated inhibition constant K(I) was 1.11.10(-8) M. The dynamic range for sulcotrione extended from 5.10(-10) M to 5.10(-6) M and the limit of detection (LOD), estimated as the concentration inducing 20% of inhibition, was 1.4.10(-10) M.

关键词