Creatinine Deiminase Adsorption onto Silicalite-Modified pH-FET for Creation of New Creatinine-Sensitive Biosensor.

Nanoscale research letters Marchenko SV, Soldatkin OO, Kasap BO, Kurc BA, Soldatkin AP, Dzyadevych SV
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In the work, silicalite particles were used for the surface modification of pH-sensitive field-effect transistors (pH-FETs) with the purpose of developing new creatinine-sensitive biosensor. Creatinine deiminase (CD) adsorbed on the surface of silicalite-coated pH-FET served as a bioselective membrane. The biosensor based on CD immobilized in glutaraldehyde vapor (GA) was taken as control. The creatinine-sensitive biosensor obtained by adsorption on silicalite was shown to have better analytical characteristics (two- to threefold increased sensitivity to creatinine, three- to fourfold lesser response and recovery times, a decrease of the detection limit of creatinine determination to 5 μM, etc.).Additionally, the biosensor based on CD adsorbed on silicalite (Sil/CD) was characterized by high signal reproducibility (relative standard deviation (RSD) for creatinine measurement = 2.6 %) and stability during storage (over 13 months). It was experimentally confirmed that the proposed biosensor was not sensitive either to high concentrations of sodium chloride or to the macromolecular protein fractions and can be used for direct quantitative analysis of creatinine in the blood serum.It was concluded that the method of CD adsorption on silicalite is well-suited for the creation of creatinine-sensitive biosensor with improved working characteristics.

关键词