Detection system based on magnetoelastic sensor for classical swine fever virus.

Biosensors & bioelectronics Guo X, Gao S, Sang S, Jian A, Duan Q, Ji J, Zhang W
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this paper, a magnetoelastic (ME) sensing system for the detection of classical swine fever virus (CSFV) is presented. The detection system comprises a test paper and a measurement circuit. The test paper consists mainly of an ME biosensor to detect the CSFV. Based on the impedance analysis technique, the measurement circuit is designed to measure the resonance frequency of the ME biosensor. The anti-CSFV IgG is immobilized onto the ME sensor surface to form the ME biosensor through a physical absorption approach. The experimental results show that the shift in the resonance frequency of the ME biosensor increases with the augmentation of the CSFV concentration. The effectiveness of the combination between the anti-CSFV IgG and CSFV is confirmed by the scanning electron microscopy (SEM) images, the sandwich-based enzyme-linked immunosorbent assay (ELISA) analysis, the interference study and the reference biosensor test method. The resonance frequency shift is linearly proportional to the concentration in the range from 0 to 2.5μg/ml, and becomes sub-linear at higher concentrations. The ME biosensor for CSFV detection has a sensitivity of about 95Hz/μg·ml(-1), with a detection limit of 0.6μg/ml.