Detection of influenza virus by a biosensor based on the method combining electrochemiluminescence on binary SAMs modified Au electrode with an immunoliposome encapsulating Ru (II) complex.

Analytical and bioanalytical chemistry Katayama Y, Ohgi T, Mitoma Y, Hifumi E, Egashira N
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Recently, point of care testing (POCT) used for diagnosis of influenza infection has a problem showing false negative diagnosis because of the low sensitivity. We would like to report detection of influenza virus A (H1N1) by an immunosensor based on electrochemiluminescence (ECL) that uses an immunoliposome encapsulating tris(2,2'-bipyridyl)ruthenium(II) complex. By using the sensor, we could detect the virus that competed with hemagglutinin (HA) peptide immobilized on self-assembled monolayers (SAMs) in immunoreaction of the antibody bound on the surface of liposome. The HA peptide was 19 mer (TGLRNGITNKVNSVIEKAA). We demonstrated great improvement of sensitivity and accuracy by introducing binary SAMs instead of mono SAMs. The binary SAMs was prepared from 3,3'-dithiodipropionic acid and 1-hexanethiol. Use of the binary SAMs enabled to increase the SAMs coverage on Au electrode; the fact was confirmed by observation of the cathodic desorption currents. By using such an electrode, first the detection method of BSA was optimized to lower ECL background signal. Then we applied the method to the detection of influenza virus. We could successfully detect the virus with higher sensitivity compared with that by POCT and ELISA. The detection range was from a concentration of 2.7 × 10(2) to 2.7 × 10(3) PFU/mL.