2016

Reduced graphene oxide-nickel nanoparticles/biopolymer composite films for the sub-millimolar detection of glucose.

The Analyst Krishna R, Campiña JM, Fernandes PM, Ventura J, Titus E, Silva AF
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Hybrid conjugates of graphene with metallic/semiconducting nanostructures can improve the sensitivity of electrochemical sensors due to their combination of well-balanced electrical/electrocatalytic properties and superior surface-to-volume ratio. In this study, the synthesis and physical characterization of a hybrid conjugate of reduced graphene oxide and nickel nanoparticles (rGO-Ni NPs) is presented. The conjugate was further deposited onto a glassy carbon electrode as a nanocomposite film of chitosan and glucose oxidase. The electrochemical response and morphology of the films were investigated using SEM, CV, and EIS, and their applications as a glucose biosensor explored for the first time in proof-of-concept tests. The low operating potential along with the good linearity and sensitivity (up to 129 μA cm(-2) mM(-1)) found in the sub-millimolar range suggest potential applications in the self-management of hypoglycemia from blood samples or in the development of non-invasive assays for body fluids such as saliva, tears or breath.

关键词