We have developed a biosensing platform for lactate determination based on gold electrodes modified with diamond nanoparticles of 4nm of nominal diameter, employing the enzyme lactate oxidase and (hydroxymethyl)ferrocene (HMF) as redox mediator in solution. This system displays a response towards lactate that is completely different to those typically observed for lactate biosensors based on other nanomaterials, such as graphene, carbon nanotubes, gold nanoparticles or even diamond nanoparticles of greater size. We have observed by cyclic voltammetry that, under certain experimental conditions, an irreversible wave (E(0)=+0.15V) appears concomitantly with the typical Fe(II)/Fe(III) peaks (E(0)=+0.30V) of HMF. In this case, the biosensor response to lactate shows simultaneous electrocatalytic peaks at +0.15V and +0.30V, indicating the concurrence of different feedback mechanisms. The achievement of a biosensor response to lactate at +0.15V is very convenient in order to avoid potential interferences. The developed biosensor presents a linear concentration range from 0.02mM to 1.2mM, a sensitivity of 6.1μAmM(-1), a detection limit of 5.3μM and excellent stability. These analytical properties compare well with those obtained for other lactate-based biosensors that also include nanomaterials and employ HMF as redox mediator.