Multifunctional Flow-injection Biosensor for the Simultaneous Measurement of Creatinine, Glucose, and Urea.

Bioscience, biotechnology, and biochemistry Rui CS, Ogawa HI, Sonomoto K, Kato Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

An amperometric flow-injection biosensor system was developed for the simultaneous measurement of creatinine, glucose, and urea, with a single sample injection and one detector. The principle for the amperometric detection of urea and creatinine was based on coupled reactions of three sequentially aligned enzyme reactors, urease or creatinine deiminase/glutamate dehydrogenase/L-glutamate oxidase. Inclusion of a split point and two confluence points between the injector and detector (oxygen electrode) resulted in a flow-injection system composed of three channels. Triple peak recording was obtained by putting two delay coils of different lengths at the channels involving urease and glucose oxidase. The system gave linear calibrations for creatinine, glucose, and urea in the ranges of 0.2-5, 0.2-10, and 0.5-20mM, respectively. The assay procedure was simple and one run was completed within 4 min. The system was reproducible within 5-8% of relative standard deviation.