2016

Dataset of the binding kinetic rate constants of anti-PCSK9 antibodies obtained using the Biacore T100, ProteOn XPR36, Octet RED384, and IBIS MX96 biosensor platforms.

Data in brief Yang D, Singh A, Wu H, Kroe-Barrett R
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Here we provide data from a head-to-head comparison study using four biosensor platforms: GE Healthcare׳s Biacore T100, Bio-Rad׳s ProteOn XPR36, ForteBio׳s Octet RED384, and Wasatch Microfluidics׳s IBIS MX96. We used these instruments to analyze the binding interactions of a panel of ten high-affinity monoclonal antibodies with their antigen, human proprotein convertase subtilisin kexin type 9 (PCSK9). For each instrument, binding curves obtained at multiple densities of surface antibodies were fit to the 1:1 Langmuir kinetic model, and the association and dissociation rate constants and corresponding affinity constants were calculated. The data supplied in this article accompany the research article entitled, "Comparison of biosensor platforms in the evaluation of high affinity antibody-antigen binding kinetics" (Yang et al., 2016) [1], which further discusses the strengths and weaknesses of each biosensor platform with an emphasis on data consistency, comparability, and operational efficiency.