2016

Surface plasmon resonance-based immunoassay for procalcitonin.

Analytica chimica acta Vashist SK, Schneider EM, Barth E, Luong JH
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A surface plasmon resonance (SPR) biosensor has been developed for rapid immunoassay of procalcitonin (PCT) with high detection sensitivity and reproducibility. The 1-ethyl-3-(3-dimethylaminopropyl)-carbodiimide hydrochloride (EDC)-activated protein A (PrA), diluted in 1% (v/v) 3-aminopropyltriethoxysilane (APTES) was dispensed on a KOH-treated Au-coated SPR chip, resulting in the covalent binding of PrA in 30 min. This "single-step" PrA immobilization strategy led to the oriented binding of the anti-PCT antibody (Ab) on a PrA-functionalized gold (Au) chip. The leach-proof immobilization procedure is five-fold faster than conventional counterparts, enabling high detection specificity and reproducibility. The IA detects 4-324 ng mL(-1) of PCT with a limit of detection (LOD) and a limit of quantification (LOQ) of 4.2 ng mL(-1) and 9.2 ng mL(-1), respectively. It was capable of detecting PCT in real sample matrices and patient samples with high precision. The Ab-bound SPR chips were stable for more than five weeks.

关键词