This study presents a novel approach based on a four-electrode electrochemical biosensor for the detection of tau protein - one of the possible markers for the prediction of Alzheimer's disease (AD). The biosensor is based on the formation of stable antibody-antigen complexes on gold microband electrodes covered with a layer of a self-assembled monolayer and protein G. Antibodies were immobilized on the gold electrode surface in an optimal orientation by protein G interaction. Electrochemical impedance spectroscopy was used to analyze impedance change, which revealed a linear response with increasing tau concentrations. The assay is fast (<1h for incubation and measurement) and very sensitive. The limit of quantification for the full-length 2N4R tau protein is 0.03pM, a value unaltered when the assay was processed in bovine serum albumin or human serum. This technology could be adapted for the detection of other biomarkers to provide a multiple assay to identify AD progression in a point of care setting.