2017

Multiplexed efficient on-chip sample preparation and sensitive amplification-free detection of Ebola virus.

Biosensors & bioelectronics Du K, Cai H, Park M, Wall TA, Stott MA, Alfson KJ, Griffiths A, Carrion R, Patterson JL, Hawkins AR, Schmidt H, Mathies RA
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

An automated microfluidic sample preparation multiplexer (SPM) has been developed and evaluated for Ebola virus detection. Metered air bubbles controlled by microvalves are used to improve bead-solution mixing thereby enhancing the hybridization of the target Ebola virus RNA with capture probes bound to the beads. The method uses thermally stable 4-formyl benzamide functionalized (4FB) magnetic beads rather than streptavidin coated beads with a high density of capture probes to improve the target capture efficiency. Exploiting an on-chip concentration protocol in the SPM and the single molecule detection capability of the antiresonant reflecting optical waveguide (ARROW) biosensor chip, a detection limit of 0.021pfu/mL for clinical samples is achieved without target amplification. This RNA target capture efficiency is two orders of magnitude higher than previous results using streptavidin beads and the limit of detection (LOD) improves 10×. The wide dynamic range of this technique covers the whole clinically applicable concentration range. In addition, the current sample preparation time is ~1h which is eight times faster than previous work. This multiplexed, miniaturized sample preparation microdevice establishes a key technology that intended to develop next generation point-of-care (POC) detection system.