2017

Research on a New Micro-Volume Fluorescence Capillary Biosensor Assay for Sequentially Quantifying Pyruvate and Lactate.

Journal of fluorescence Li YS, Li QJ, Yang W, Gao XF
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

It was studied that making conditions of a micro-volume fluorescence capillary biosensor for determining pyruvate (PA) and lactate (LA). The biosensor made under the optimized conditions could be used for sequential quantifications of LA in the range 0.10-1.2 mM and PA in 4-120 μM, and its recovery for PA and LA was in a satisfactory range 97-106% for human serum samples, with detection limits of 0.023 mM for LA (RSD < 1.89%, n = 11) and 0.87 μM for PA (RSD < 1.70%, n = 11). The new assay possessed these advantages that the LDH immobilizing on capillary realized the reuse of expensive enzyme in fluorospectrophotometry, and the consumption of serum samples or chemical reagents decreased to 9 μL in per assay, and the analytes no needed to preseparation, and it also are accurate and reliable. Consequently, the fluorescence capillary biosensor should have a good prospect in assaying PA and LA or LA/PA ratios for clinical medicines or biology field. The optimization conditions and parameters obtained in this study have also a certain guiding significance for the development of biochip based on glass substrate.

关键词