2017

Human heart failure biomarker immunosensor based on excessively tilted fiber gratings.

Biomedical optics express Luo B, Wu S, Zhang Z, Zou W, Shi S, Zhao M, Zhong N, Liu Y, Zou X, Wang L, Chai W, Hu C, Zhang L
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A label-free immunosensor platform based on excessively tilted fiber gratings (Ex-TFGs) was developed for highly specific and fast detection of human N-terminal pro-B-type natriuretic peptide (NT-proBNP), which is considered a powerful biomarker for prognosis and risk stratification of heart failure (HF). High-purity anti-NT-proBNP monoclonal antibodies (MAbs) prepared in our laboratory were immobilized on fiber surface through the staphylococcal protein A (SPA) method for subsequent specific binding of the targeted NT-proBNP. Utilizing fiber optic grating demodulation system (FOGDS), immunoassays were carried out in vitro by monitoring the resonance wavelength shift of Ex-TFG biosensor with immobilized anti-NT-proBNP MAbs. Lowest detectable concentration of ~0.5ng/mL for NT-proBNP was obtained, and average sensitivity for NT-proBNP at a concentration range of 0~1.0 ng/mL was approximately 45.967 pm/(ng/mL). Several human serum samples were assessed by the proposed Ex-TFG biomarker sensor, with high specificity for NT-proBNP, indicating potential application in early diagnosing patients with acute HF symptoms.

关键词