2017

DNA-wrapped multi-walled carbon nanotube modified electrochemical biosensor for the detection of Escherichia coli from real samples.

Talanta Ozkan-Ariksoysal D, Kayran YU, Yilmaz FF, Ciucu AA, David IG, David V, Hosgor-Limoncu M, Ozsoz M
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This paper introduces DNA-wrapped multi-walled carbon nanotube (MWCNT)-modified genosensor for the detection of Escherichia coli (E. coli) from polymerase chain reaction (PCR)-amplified real samples while Staphylococcus aureus (S. aureus) was used to investigate the selectivity of the biosensor. The capture probe specifically recognizing E. coli DNA and it was firstly interacted with MWCNTs for wrapping of single-stranded DNA (ssDNA) onto the nanomaterial. DNA-wrapped MWCNTs were then immobilised on the surface of disposable pencil graphite electrode (PGE) for the detection of DNA hybridization. Electrochemical behaviors of the modified PGEs were investigated using Raman spectroscopy and differential pulse voltammetry (DPV). The sequence selective DNA hybridization was determined and evaluated by changes in the intrinsic guanine oxidation signal at about 1.0V by DPV. Numerous factors affecting the hybridization were optimized such as target concentration, hybridization time, etc. The designed DNA sensor can well detect E. coli DNA in 20min detection time with 0.5pmole of detection limit in 30µL of sample volume.

关键词