Simulations and design of microfabricated interdigitated electrodes for use in a gold nanoparticle enhanced biosensor.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference Hermansen P, MacKay S, Wishart D, Jie Chen
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Microfabricated interdigitated electrode chips have been designed for use in a unique gold-nanoparticle based biosensor system. The use of these electrodes will allow for simple, accurate, inexpensive, and portable biosensing, with potential applications in diagnostics, medical research, and environmental testing. To determine the optimal design for these electrodes, finite element analysis simulations were carried out using COMSOL Multiphysics software. The results of these simulations determined some of the optimal design parameters for microfabricating interdigitated electrodes as well as predicting the effects of different electrode materials. Finally, based on the results of these simulations two different kinds of interdigitated electrode chips were made using photolithography.