2016

A wide range and high sensitivity four-channel compact electrochemical biosensor for neurotransmitter detection on a microfluidic platform.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference Ghodsevali E, Landari H, Boukadoum M, Gosselin B, Miled A
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We present a four-channel, high-sensitivity and linearity electrochemical biosensor for neurotransmitter (NT) detection and measurement. Using a multi-channel microfluidic platform makes this biosensor capable of detecting NT-related currents going from nanoamperes to milliamperes, with a sensitivity of the order of picoamperes. Moreover, by using a fully differential potentiostat architecture, the biosensor offers a high common-mode rejection ratio (90 dB), making it appropriate for low-noise and high-sensitive applications. The system was implemented on a 15 mm × 15 mm PCB with direct interface to the microfluidic chambers. It was calibrated with a 5 mM ferrocyanide solution and successfully tested with dopamine at three concentrations. The system shows a minimum sensistivity of 100 pA and consumes 60 mW.

关键词